Specificaties

 

ECG General

  • Full 12-Lead ECG with 9 independent outputs for each signal lead referenced
    to RL.
  • Output Impedances: 500, 1000, 1500, & 2000 ohms to RL.
  • High Level Output: 0.5 V/mV of low level selection.
  • Amplitude Accuracy: ± 2% 2 Hz Square Wave (Lead II).

Normal Sinus Rhythm

  • Rates: 30, 40, 60, 80, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, 300
  • BPM. Accuracy ± 1%.
  • Amplitudes (Lead II): 5mV, 4mV, 3mV, 2mV, 1mV, .5mV,.25mV,.1mV.
  • ST Segments: 16 total – 8 elevated & 8 depressed.
  • ST Segment Levels (Lead II): -0.8 mV to +0.8 mV in 0.1 mV steps on Lead II, .
  • Axis Deviation: Normal (intermediate), horizontal, and vertical. Modifies
    baseline ECG during arrhythmias.
  • Neonatal Mode: ECG R wave width is reduced to 40ms.

ECG Performance Testing

  • Square Wave: 2 Hz
  • Square Wave: 0.125 Hz
  • Pulse: 4.0 secs
  • Sine Waves: 0.05, 0.5, 1, 10, 25, 30, 40, 50, 60, and 100 Hz.
  • Triangle Wave: 2 Hz
  • R Wave Detector Test: 60 BPM haver-triangle wave with selectable ampl. and
    width.
  • Width: 8.0 ms to 200 ms (12 selections)
  • Amplitude (Lead II and V Leads): 5mV to 0.5mV.

Pacemaker

  • Asynchronous
  • Demand with frequent sinus beat
  • Demand with occasional sinus beat
  • A-V sequential
  • Non-capture non-function
  • Rhythms: Async 75 BPM, Demand 1, Demand 2, AV Seq, Non capture,
    Non function. - Pulse: -700 mV to +700 mV. Accuracy : 10%
  • Width: 0.1, 0.2, 0.5, 1.0, 2.0 ms. Accuracy is 5%.
  • Pulse Polarity: Positive or negative.

Synchronization

  • Accepts input during defib training.

RS-232 Interface

  • RS-232 interface to PC.

Defibrillator Training

  • With two emergency scenarios and a cardioversion procedure, AMPS-1 can
    be used for basic defibrillator training.

Cardiac Output

  • Built in Cardiac Output feature, activation optional
  • 4 adjustable injectate temperature selections that are factory set and user
    adjustable
  • Baseline of 36, 37 and 38 degrees °C.
  • Selections for 2 and 20 degrees °C.

Cardiac Output Selections

  • Faulty Injectate Curve
  • Left to Right Shunt Curve
  • C.O. of 3, 4, 4.5, 5, 5.5, 6, 6.5, 7 l/min
  • Cal Pulse: 1o for 1 second
  • Cal Pulse: 1o for 4 seconds
  • CC .561 for 2 degrees injectate
  • CC .608 for 20 degrees injectate

Temperature

  • 2 Temperature Channels
  • Electronically Switched Temperature of 35, 37, 38, 40, 42°C.
  • Probe Compatibility: 400 or 700 series YSI

Respiration

  • Baseline Impedance: 500, 1000, 1500, 2000 ohms, LEADS I, II, III
  • Impedance Variations: 5, 4, 3, 2, 1, 0.5, 0.2, 0.1, 0.05, 0 O
  • Rates: 15 to 120 and 0 rpm for APNEA
  • Apnea Selections: 12, 22, 32 seconds, and continuous
  • Respiratory Effort (Inspiration/Expiration Ratio:) 5/1, 4/1, 3/1 (normal), 2/1, 1/1.

Optional Fetal / Maternal / IUP Simulations

  • Fetal heart rates:60, 90, 120, 140, 150, 210, & 240 BPM
  • Trend
  • Uniform, Early and Late deceleration
  • Maternal heart rate fixed at 80 BPM
  • Waveform:12 lead ECG with complete p-qrs-t complex
  • Dynamic intrauterine pressure (iup) waveform:positive bell shaped pressure curve
  • Peak pressure:90 mmhg, Contraction duration:90 sec
  • Pressure transducer sensitivity:5 or 40 m v/v/mmhg
  • Input/output impedance:300 ohms
  • Optional Mechanical Fetal Heart

Arrhythmia Selections

  • Premature BeatsPremature Atrial Contraction (PAC)
  • Nodal Premature Nodal Contraction (PNC)
  • Premature Ventricular Contraction (PVC)1
  • Left Ventricular Focus
  • PVC1 Early, Left Ventrical (LV) Focus PVC1 R-on-T, Left Ventrical (LV) Focus
  • PVC2 Right Ventricular Focus PVC2 Early, RV Focus
  • PVC2 R-on-T RV Focus Multifocal PVCs
  • AED test waveforms
    • Atrial Fibrillation, Course Atrial fibrillation, Fine
    • Ventricular Fibrillation, Course Ventricular Fibrillation, Fine
    • Asystole, Flatline
    • Supraventricular Tachycardia Ventricular Tachycardia @140,160 & 190BPM
    • Torsades de Pointes @ 200 BPM NSR @ 60 BPM
  • Conduction Defects
  • First Degree Heart Block Second Degree Heart Block
  • Third Degree Heart Block Right Bundle Branch Block
  • Left Bundle Branch Block Supraventricular Beats
  • Atrial Fibrillation (Coarse) Atrial Fibrillation (Fine)
  • Atrial Flutter Sinus Arrhythmia
  • Missed Beat (1 time event) Paroxysmal Atrial Tachycardia
  • Nodal Supraventricular Tachycardia
  • Ventricular Rhythm PVCs 6/Minute
  • PVCs 12/Minute PVCs 24/Minute
  • Frequent Multifocal Asystole
  • Pair PVCs (1 time event) Run 5 PVCs (1 time event)
  • Run 11 PVCs (1 time event) Ventricular Tachycardia
  • Ventricular Fibrillation (Coarse) Ventricular Fibrillation (Fine)
  • Bigeminy Trigeminy

4 Blood Pressure Channels

  • Electrically Isolated Channels
  • Dynamic BP waveforms are synchronized with normal sinus rhythm rates and track arrhythmia selections.
  • Respiration artifact can be selected on blood pressure channels
  • Transducer Sensitivity: 5 or 40 µV/V/mmHg
  • Calibrated Rate: 80 BPM normal sinus rhythm
  • Static Levels BP1/2/3/4: -10, -5, 0, 20, 40, 50, 60, 80, 100, 120, 150, 160, 180, 200,
    240, 320, 400 mmHg
  • Automatic Swan-Ganz (every 15 seconds)
  • Manual Swan-Ganz, changes each time Enter is selected
 